Patients
Eclipse FAQs
|
|
Inches |
Centimeters |
|
Height |
19.3 |
49.0 |
|
Width |
12.3 |
31.2 |
|
Depth |
7.1 |
18.0 |
- Without power cartridge: 14.5 pounds (6.6 kg)
|
|
Typical New Power Cartridge Duration-Of-Use (Battery) | |
|
FLOW SETTING |
Continuous Flow Mode |
Pulse Flow Mode |
|
0.5 |
4.4 hours |
- |
|
1.0 |
3.7 hours |
4.4 hours |
|
2.0 |
1.8 hours |
3.6 hours |
|
3.0 |
1.3 hours |
3.0 hours |
|
4.0 |
- |
2.6 hours |
|
5.0 |
- |
2.3 hours |
|
6.0 |
- |
1.8 hours |
Note: Times will degrade with Power Cartridge use, over time, higher bolus size, breath rate and the increasing age of the Power Cartridge.
AutoSat is a robust pulse delivery system that maintains a consistent bolus size for every breath as your patients’ respiratory rate increases or decreases. Therefore, if the patient has been properly titrated, the clinical efficacy of bolus size and FIO2 are never compromised.
Many airlines have approved the Eclipse for use on board their airlines. Each airline may have specific requirements for patients traveling with their own portable oxygen system. The patient should check with the airline for details on approvals and special arrangements to bring the Eclipse on the airplane.
Click here for more information.
- Continuous flow – up to 50ft.
Check the air inlet filter daily and clean it at least once a week.
- Remove the filter.
- Wash the filter in warm water using a mild detergent solution.
- Rinse the filter thoroughly and allow it to dry.
- When completely dry, reinsert the filter into the cabinet.