Eclipse 3 Specifications
It is important to differentiate between the pulse dose only units on the market and the Eclipse, a true Personal Ambulatory Oxygen System (PAOS) offers both continuous flow and pulse dose options. Pulse dose does not have an accepted continuous flow equivalent. Having both continuous flow and pulse dose allows patients the ability to use one unit for a variety of activities. Proper titration in both modes remains an important component for patients to consider. Patients need to be confident they will be properly oxygenated during activities of daily living, sleep and at altitude.
Eclipse® 3 with autoSAT™ Technology
Personal Ambulatory Oxygen System™ (PAOS)
One simple and easy-to-use unit... with the Eclipse 3 you are no longer afraid of running out of oxygen.
Features and Benefits
· Continuous flow up to 3LPM and a pulse dose option
- Continuous flow is the gold standard for nocturnal use
· autoSAT™ Technology adjusts to your breath rate to ensure consistent bolus volume
· AC/DC and power cartridge allows for travel around town or even around the world
· Reduced power consumption means lower electrical costs
· FAA Approved
· Operates up to 13,123 feet in altitude for air or land excursions
· Is compatible with many leading CPAP and bi-level devices in Continuous Flow Mode
The Eclipse 3 comes complete with a Power Cartridge, AC Power Suppy, DC Power Supply, Universal Cat with Telescoping handle, cannula, as well as, an informative Patient Pack.
|
Product Specifications |
|
|
Size (H x W x D):
|
19.3 x 12.3 x 7.1 inches |
|
49.0 x 31.2 x 18.0 cm |
|
|
Weight: |
Eclipse: 15.0 lbs |
|
Power Cartridge: 3.4 lbs |
|
|
Continuous flow settings Pulse Dose Settings (measured in mL) |
0.5 to 3.0 LPM (0.5 liter increments) 16-96 (16mL increments), 128mL, 160mL, 192mL |
|
Oxygen concentration |
90% ± 3% @ sea level |
|
Oxygen output pressure |
5.0 psig (34.5 kPa) nominal |
|
Sound level |
48 dB(A) at 3.0 LPM continuous flow |
|
40 dB(A) at 3.0 Pulse Dose setting |
|
|
Power specifications: |
|
|
AC Operation |
100 - 240 VAC, 50 - 60 Hz |
|
DC Operation |
12 V nominal |
|
Power Cartridge |
Quantity (2) 97.5 W-Hr battery packs each containing |
|
7.92 grams of equivalent lithium content |
|
|
Power Consumption |
52 watts at 1.0 LPM continuous flow |
|
145 watts at 3.0 LPM continuous flow |
|
|
45 watts at 1.0 pulse flow setting |
|
|
95 watts at 6.0 pulse flow setting |
|
|
Nominal Power Cartridge operating time |
2.0 hours at 2.0 LPM continuous flow |
|
1.3 hours at 3.0 LPM continuous flow |
|
|
5.2 hours at 2.0 pulse flow setting, 12 bpm |
|
|
3.5 hours at 6.0 pulse flow setting, 12 bpm |
|
|
Power Cartridge recharge time |
1.8 to 5.0 hrs, dependent upon the flow setting, to achieve 80% capacity from a fully discharged Power Cartridge |
|
Back up alarm power |
9 Volt battery |
|
Oxygen concentration indicator |
Green Light =
|
|
Yellow Light = Warning or Caution |
|
|
Red Light Flashing = Abnormal Operation |
|
|
Operating temperature |
50° F to 104° F (10° C to 40° C) |
|
|
82.4° F (28° C) dew point, non-condensing |
|
Trigger sensitivity |
Adjustable between -0.25 cm H2O to -0.7 cm H2O |
|
Altitude operating range |
0-13,123 feet (0-4000 meters) |
* Specifications subject to change without notice